Mohamed M. Hefnawy, Maha A. Sultan and Mona M. Al-Shehri
Department of Pharmaceutical Chemistry, College of Pharmacy, King Saud University, P.O. Box 2457, Riyadh 11451, Saudi Arabia.
Abstract: A high-performance liquid chromatographic (HPLC) method has been developed for the separation and determination of S- and R-enantiomers of betaxolol in tablets and ophthalmic preparations. Baseline resolution was achieved by using teicoplanin macrocyclic antibiotic chiral stationary phase (CSP) known as Chirobiotic T with fl uorescence detection at excitation/emission wavelengths 275/305 nm. The polar ionic mobile phase (PIM) consists of methanol-glacial acetic acid-triethylamine, (100:0.020:0.025, v/v/v) has been used at a fl ow rate of 1.5 ml/min. All analytes with S-(-)-atenolol as internal standard were conducted at ambient temperature. The method is highly specific where another coformulated compounds did not interfere. The stability of betaxolol enantiomers under different degree of temperature also studied. The results showed that it is stable for at least 7 days at 70oC. The method validated for its linearity, accuracy, precision and robustness. Experimental design was used during validation to evaluate method robustness. Using the chromatographic conditions described, S- and R-betaxolol were well resolved with mean retention times of 11.3 and 12.6 min, respectively.
Linear response (r > 0.997) was observed over the range of 10-500 ng/ml of betaxolol enantiomers, with detection limit of 5 ng/ml. The recoveries of S- and R-betaxolol from tablets and ophthalmic preparation ranged from 97.4 to 101.4% and 98.0 to 102.0%, respectively. The mean relative standard deviation (R.S.D.%) for both enantiomers were 1.1-1.4% and 1.3-1.7% in tablets and ophthalmic solution, respectively.
Discussion
No comments yet...Be the first to comment.
The publication of our paper in Bioinformatics and Biology Insights was highly professional and very pleasant on all levels: the guidelines for authors are concise, the online submission system is user-friendly, the comments from the reviewers were insightful and improved our paper, and the preparation of the manuscript for publication was efficient. I particularly liked the fast feedback from the staff on the state of the submission and review process.Professor Werner Braun (Department of Biochemistry and Molecular Biology, Sealy Center for Structural Biology and Molecular Biophysics, University of Texas Medical Branch, Galveston, Texas, USA) What our authors say
Copyright © 2011 Libertas Academica Ltd (except open access articles and accompanying metadata and supplementary files.)