Clinical Medicine Insights: Cardiology 2015:9 85-90

Original Research

Published on 12 Aug 2015

DOI: 10.4137/CMC.S24976


Further metadata provided in PDF



Sign up for email alerts to receive notifications of new articles published in Clinical Medicine Insights: Cardiology

Abstract

Background: The impact of magnetic resonance imaging (MRI) on pacemakers is potentially hazardous. We present clinical results from a novel MRI conditional pacing system with the capability to switch automatically to asynchronous mode in the presence of a strong magnetic field.

Aims: The IKONE (Assessment of the MRI solution: KORA 100™ and Beflex™ pacing leads system) study is an open-label, prospective, multicenter study aimed at confirming the safety and effectiveness of the system, when used in patients undergoing MRI of anatomical regions excluding the chest.

Methods: Primary eligibility criteria included patients implanted with the system, with or without a clinically indicated MRI. The primary endpoint was to confirm no significant change in pacing capture thresholds at 1 month after an MRI, with an absolute difference of ≤0.75 V between the pre- and 1-month post-MRI for both atrial and ventricular capture thresholds.

Results: Out of 33 patients enrolled (mean age: 72.8 ± 11.4 years, 70% male, implant indication or device), 29 patients implanted with the MRI conditional system underwent an MRI 6–8 week postimplant. The study reached its primary endpoint: the mean absolute difference in pacing capture threshold at 1-month post-MRI versus pre-MRI was less than 0.75 V in the atrium (Δ = 0.18 ± 0.16 V, P-value <0.001) and in the ventricle (Δ = 0.18 ± 0.22 V, P-value <0.001). There were no adverse events related to the MRI procedure nor were there reports of patient symptoms or discomfort associated. MR image quality was of diagnostic quality in all patients.

Conclusion: Lead electrical performance as measured by difference in capture thresholds were not impacted by MRI. This first clinical evaluation of a novel MRI conditional system demonstrates it is safe and effective for use in out-of-chest, 1.5-T MR imaging.



Downloads

PDF  (413.59 KB PDF FORMAT)

RIS citation   (ENDNOTE, REFERENCE MANAGER, PROCITE, REFWORKS)

BibTex citation   (BIBDESK, LATEX)


Sharing


What Your Colleagues Say About Clinical Medicine Insights: Cardiology
The review process in Libertas Academica journals is prompt, objective and rigorous.  The entire process is smooth and clear and all the editorial staff are cooperative and professional.  Published papers are of top clinical and scientific value, and high metrics reflect the excellent quality of Libertas Academica's journals.
Dr Antoine Kossaify (USEK NDS University Hospital, Byblos, Lebanon)
More Testimonials

Quick Links


New article and journal news notification services
Email Alerts RSS Feeds
Facebook Google+ Twitter
Pinterest Tumblr YouTube