Publication Date: 02 Dec 2013
Type: Case report
Journal: Clinical Medicine Insights: Case Reports
Citation: Clinical Medicine Insights: Case Reports 2013:6 183-187
doi: 10.4137/CCRep.S13380
We report on a 70-year-old male patient who was recipient of GEM III DR 7275 Cardioverter Defibrillator, and who presented with inappropriate shocks. The patient had a documented slow ventricular tachycardia (VT), and the device was programmed to detect VT at rates >100 bpm, fast VT (FVT, via VT) at rates >150 bpm, and ventricular fibrillation (VF) at rates >188 bpm. After detection of FVT, efficient therapy was delivered; however, this was immediately followed by multiple inappropriate therapies. Inappropriate therapies were discussed, with a focus on programming features.
PDF (2.95 MB PDF FORMAT)
RIS citation (ENDNOTE, REFERENCE MANAGER, PROCITE, REFWORKS)
BibTex citation (BIBDESK, LATEX)
PMC HTML
I found Clinical Medicine Insights: Case Reports very, very user friendly. The entire process is easy and straightforward. The corresponding author is kept updated on the progress at every point. I am pleased to send this endorsement.
Facebook Google+ Twitter
Pinterest Tumblr YouTube