We report on a 70-year-old male patient who was recipient of GEM III DR 7275 Cardioverter Defibrillator, and who presented with inappropriate shocks. The patient had a documented slow ventricular tachycardia (VT), and the device was programmed to detect VT at rates >100 bpm, fast VT (FVT, via VT) at rates >150 bpm, and ventricular fibrillation (VF) at rates >188 bpm. After detection of FVT, efficient therapy was delivered; however, this was immediately followed by multiple inappropriate therapies. Inappropriate therapies were discussed, with a focus on programming features.


Downloads

PDF  (2.95 MB PDF FORMAT)

RIS citation   (ENDNOTE, REFERENCE MANAGER, PROCITE, REFWORKS)

BibTex citation   (BIBDESK, LATEX)

XML

PMC HTML


Sharing




What Your Colleagues Say About Clinical Medicine Insights: Case Reports
I found Clinical Medicine Insights: Case Reports very, very user friendly.  The entire process is easy and straightforward.  The corresponding author is kept updated on the progress at every point.  I am pleased to send this endorsement.
Dr Adebayo Oyedeji (Lekki Hospital, Nigeria)
More Testimonials

Quick Links


New article and journal news notification services
Email Alerts RSS Feeds
Facebook Google+ Twitter
Pinterest Tumblr YouTube